Get upto 25% OFF on All Order - SHOP NOW

The CONTEC CA10S Capnograph is a compact, handheld patient monitor that adopts the sidestream sampling method to continuously measure End-Tidal Carbon Dioxide ($\text{EtCO}_2$) and Airway Respiration Rate ($\text{AWRR}$). It is designed to evaluate pulmonary ventilation functions, reflect circulation and pulmonary blood flow conditions, and indirectly track alveolar ventilation. Concurrently, it features integrated pulse oximetry to monitor blood oxygen saturation ($\text{SpO}_2$) and pulse rate via a dedicated sensor probe.
| Parameter / Feature | Specification Details |
| $\text{EtCO}_2$ Measurement Range | 0 to 150 mmHg |
| $\text{EtCO}_2$ Resolution | 1 mmHg |
| $\text{EtCO}_2$ Accuracy |
0–40 mmHg: $\pm$2 mmHg 41–70 mmHg: $\pm$5% of reading 71–100 mmHg: $\pm$8% of reading 101–150 mmHg: $\pm$10% of reading |
| Respiration Rate Range | 2 rpm to 120 rpm (breaths per minute) |
| Respiration Accuracy | $\pm$1 rpm |
| $\text{SpO}_2$ Measurement Range | 30% to 100% (Resolution: 1%) |
| $\text{SpO}_2$ Accuracy | 70% to 100%: $\pm$2% (0% to 69%: Undefined) |
| Pulse Rate Range | 25 bpm to 250 bpm (Resolution: 1 bpm) |
| Pulse Rate Accuracy | $\pm$2 bpm or $\pm$2% (whichever is greater) |
| Power Supply | DC 3.7V Rechargeable Lithium Battery |
| Input Power | $\le$1.5 VA |
| Physical Dimensions | 159 mm $\times$ 72 mm $\times$ 31 mm |
| Device Weight | 270 g |
| Safety / Protection Grade | Internally powered equipment, Type BF applied part, IPX1 liquid ingress protection |
Multi-Parameter Monitoring: Seamless tracking of End-tidal $\text{CO}_2$, Respiratory Rate ($\text{AWRR}$), Oxygen Saturation ($\text{SpO}_2$), and Pulse Rate ($\text{PR}$).
Advanced Safety Alarms: Built-in audio and visual over-limit alert system for all four core vital metrics, plus low battery indicators.
Intelligent Compensation: Built-in atmospheric pressure and temperature auto-compensation to ensure accurate readings regardless of environment.
Self-Calibration: Intuitive feature allowing users to perform zero-point gas calibration manually.
Maintenance Alerts: Includes a "Checking Filterline" automated notification to prompt for system line inspection.
Bright Visual Display: Equipped with a 2.8" color LCD screen displaying customizable configuration limits and unit options ($\text{mmHg}$, $\text{kPa}$, or %).
Ultra-Portable Profile: Extremely lightweight (270g) and compact ergonomic body tailored for quick transport and bedside use.
Extended Battery Endurance: The internal rechargeable lithium battery operates up to 8 hours on a full charge.
Target Environments: Pre-hospital emergency transport, Emergency Departments (ED), Intensive Care Units (ICU), operating rooms, neonatology units, and respiratory wards.
Target Patients: Universally applicable across adult, pediatric, and neonatal patient populations.
CONTEC CA10S Capnograph Main Unit
Adult/Pediatric Blood Oxygen ($\text{SpO}_2$) Probe
Sidestream Gas Sampling Pipe (Filterline)
USB Cable/Wire (for charging and data transmission)
User Manual