Get upto 25% OFF on All Order - SHOP NOW

| Patient Type | Adult & Pediatric |
| Ventilation Mode | Invasive & NIV |
| Application | Hospital |
| Tidal Volume | 20-1500 ml |
| Display Size | 7 inch |
| Power Source | Electric |
| Usage/Application | ICU |
| Type | Ventilator |
| Display Type | Digital |
| Patient Age Group | Adult |
| Voltage (V) | 100 - 240 V |
| Frequency (Hz) | 50 - 60 Hz |
| Automation Grade | Automatic |
| Color | White |
| Is It Portable | Non Portable |
| Brand | Drager |
Enriched by our vast industrial experience in this business, we are involved in offering an enormous quality range of Drager Savina Ventilator.
Patient Population: Adults, Pediatrics, and Neonates (with neonatal option)
Volume-Controlled Modes:
IPPV / CMV: Intermittent Positive Pressure Ventilation
SIMV: Synchronized Intermittent Mandatory Ventilation
Pressure-Controlled Modes:
BIPAP / PCV: Biphasic Positive Airway Pressure
Spontaneous Modes:
CPAP / ASB: Continuous Positive Airway Pressure / Assisted Spontaneous Breathing (Pressure Support)
PPS: Proportional Pressure Support (optional)
Non-Invasive Ventilation: NIV support available
| Parameter | Setting Range |
| Tidal Volume ($V_T$) | $0.02\text{ L}$ to $2.0\text{ L}$ (Adult / Pediatric) |
| Inspiratory Time ($T_i$) | $0.1\text{ s}$ to $10\text{ s}$ |
| Ventilation Frequency ($f$) | $0\text{ to } 100\text{ bpm}$ |
| PEEP / Interm. PEEP | $0\text{ to } 35\text{ mbar}$ ($\text{cmH}_2\text{O}$) |
| Inspiratory Pressure ($P_{insp}$) | $0\text{ to } 99\text{ mbar}$ |
| Pressure Support ($P_{supp}$) | $0\text{ to } 35\text{ mbar}$ |
| $\text{FiO}_2$ (Oxygen Concentration) | $21\%\text{ to } 100\%$ |
| Flow Trigger Sensitivity | $0.3\text{ to } 15\text{ L/min}$ |
Graphic Screen: Displays real-time pressure, flow, and volume curves/waveforms.
LED Numeric Displays: Quick-read LED indicators for Tidal Volume, Frequency, PEEP, and $\text{FiO}_2$.
Monitored Values:
Peak Pressure ($P_{peak}$), Mean Pressure ($P_{mean}$), PEEP
Expired Minute Volume ($MV$), Expired Tidal Volume ($V_{Te}$)
Airway Temperature and Oxygen Concentration ($\text{FiO}_2$)
Power Supply: $100\text{ to } 240\text{ V AC}$, $50/60\text{ Hz}$
Battery Backup: Internal rechargeable battery (approx. 45–60 minutes runtime)
Gas Supply Pressure: $\text{O}_2$ and Medical Air supplied at $3\text{ to } 6\text{ bar}$ ($43.5\text{ to } 87\text{ psi}$)
vender - kaarya